Biomechanical Repercussions of Bone Resorption Related to Biologic Width: A Finite Element Analysis of Three Implant-Abutment Configurations
Warning: include(/home/quintpub/public_html/journals/prdincludes/05_update/javascript.php) [function.include]: failed to open stream: No such file or directory in /home/quintpub/public_html/journals/prd/abstract.php on line 39

Warning: include() [function.include]: Failed opening '/home/quintpub/public_html/journals/prdincludes/05_update/javascript.php' for inclusion (include_path='.:/usr/lib/php:/usr/local/lib/php') in /home/quintpub/public_html/journals/prd/abstract.php on line 39
Follow Us      

LOGIN

   Official Journal of The Academy of Osseointegration

 
Share Page:
Back

Volume 29 , Issue 5
September/October 2009

Pages 479-487


Biomechanical Repercussions of Bone Resorption Related to Biologic Width: A Finite Element Analysis of Three Implant-Abutment Configurations

Xavier Rodríguez-Ciurana/Xavier Vela-Nebot, MD, DDS/Maribel Segalà-Torres, MD, DDS/Carlos Rodado-Alonso, MD/Víctor Méndez-Blanco, DDS/Marta Mata-Bugueroles


PMID: 19888491
DOI: 10.11607/prd.00.0881

Most biomechanical studies of implant-supported restorations have not taken into account the biologic changes that occur following exposure of the implants to the oral environment. Therefore, the present finite element analysis study was designed to compare the biomechanical response of three types of implant-abutment configurations both before and after establishment of a new biologic width. The three functional units studied were: a 5-mm implant platform connected with an external hexagon to a 5-mm-diameter abutment (type 1), a 5-mm implant platform connected with an external hexagon to a 4.1-mm-diameter abutment (type 2), and a 4.8-mm implant platform connected with an internal hexagon to a 4.1-mm-diameter abutment (type 3). The type 3 design, which combined platform switching with an internal connection, exhibited the smallest distortions in stress distribution after bone modeling, and the stress was distributed over the entire contact surface most smoothly and uniformly. Bone resorption following creation of the biologic width changes the biomechanical behavior of a restoration. In this study, the two implant-abutment designs featuring a smaller-diameter abutment on a larger-diameter implant platform achieved better results than the implant featuring the same-diameter implant platform and abutment, even though their initial bio­mechanical load potential was lower. (Int J Periodontics Restorative Dent 2009;29:479–487.)


Full Text PDF File | Order Article

 

 
Get Adobe Reader
Adobe Acrobat Reader is required to view PDF files. This is a free program available from the Adobe web site.
Follow the download directions on the Adobe web site to get your copy of Adobe Acrobat Reader.

 

© 2020 Quintessence Publishing Co, Inc

PRD Home
Current Issue
Ahead of Print
Archive
Author Guidelines
About
Submission Form
Submit
Reprints
Permission
Advertising
Quintessence Home
Terms of Use
Privacy Policy
About Us
Contact Us
Help